眼前节测量评估系统与光学生物测量仪在有晶状体眼人工晶状体植入术中晶状体预测的准确性分析

魏安海 邹杨成 贾梦佳 刘安发 张和华

中国医学装备2025,Vol.22Issue(8):41-45,5.
中国医学装备2025,Vol.22Issue(8):41-45,5.DOI:10.3969/j.issn.1672-8270.2025.08.008

眼前节测量评估系统与光学生物测量仪在有晶状体眼人工晶状体植入术中晶状体预测的准确性分析

Analysis on the accuracy of anterior segment measurement assessment system and optical biometer in predicting lens in ICL implantation

魏安海 1邹杨成 2贾梦佳 3刘安发 3张和华4

作者信息

  • 1. 陆军军医大学大坪医院医学工程科 重庆 400042||重庆大学微电子与通信工程学院 重庆 400044
  • 2. 陆军军医大学大坪医院眼科 重庆 400042
  • 3. 西南医科大学医学信息与工程学院 四川 646699
  • 4. 陆军军医大学大坪医院医学工程科 重庆 400042
  • 折叠

摘要

Abstract

Objective:To explore the accuracy of parameters of anterior segment measurement assessment system and optical biometer in predicting lens in implantable collamer lens(ICL)implantation.Methods:This study retrospectively analyzed the examination data of patients who underwent ICL surgery at the department of ophthalmology,Army Characteristic Medical Center from January 2020 to July 2024.The measured data of the chamber corner parameters,which included anterior chamber depth(ACD),central corneal thickness(CCT),white to white(WTW),flat keratometry(K1)and steep keratometry(K2),of 74 patients(134 eyes)who underwent simultaneous measurements of the IOLMaster 700 optical biometer and the Pentacam HR anterior segment measurement assessment system were selected.The expected spherical equivalent(EXP SPH),expected cylinder(EXP CYL),and expected spherical equivalent(EXP SEQ)were calculated,and the differences,correlations and consistencies of these data were analyzed.Results:There were significant differences in ACD,CCT,WTW,K1 and K2 between the IOLMaster 700 and Pentacam HR in measuring myopic patients(t=12.928,6.202,-37.877,-13.743,-12.007,P<0.001).All measured data showed high correlation(r=0.983,0.984,0.961,0.996,0.994,P<0.001).Bland-Altman consistency test showed that the measured values by 3%-6%were outside the limits of 95%consistency,and the differences among the EXP SPH of two instruments and the postoperative actually measured results of spherical lens(SPH),among the EXP CYL of two instruments and the postoperative actually measured results of cylinder lens(CYL),and among the EXP SEQ of two instruments and the postoperative actual measured results of spherical equivalent(SEQ)were statistically significant(x2=20.927,7.451,25.459,P<0.05).The average differences between Pentacam HR and IOLMaster 700 were respectively(0.024±0.491)D and(0.009±.491)D in predicting SPH,and were respectively(-0.05±0.476)D and(-0.049±0.475)D in predicting CYL,and were respectively(-0.001±0.458)D and(-0.016±0.461)D in predicting SEQ.Conclusion:During selection process of ICL surgery for lens,clinicians can reasonably choose Pentacam HR and IOLMaster 700 to calculate lens according to the actual requirements of the patients.

关键词

有晶状体眼人工晶状体植入术(ICL)/眼晶状体预测/房角参数/近视患者

Key words

Implantable collamer lens(ICL)implantation/Prediction for eye lens/Chamber corner parameter/Myopic patient

分类

医药卫生

引用本文复制引用

魏安海,邹杨成,贾梦佳,刘安发,张和华..眼前节测量评估系统与光学生物测量仪在有晶状体眼人工晶状体植入术中晶状体预测的准确性分析[J].中国医学装备,2025,22(8):41-45,5.

基金项目

重庆市创新医疗器械应用示范项目(CQEIC2024MDAD-018) (CQEIC2024MDAD-018)

重庆市技术创新与应用发展专项重点项目(CSTB2021TIAD-KPX0074) Chongqing Innovative Medical Devices Application Demonstration Project(CQEIC2024MDAD-018) (CSTB2021TIAD-KPX0074)

Key Projects of the Chongqing Technological Innovation and Application Development Fund(CSTB2021TIAD-KPX0074) (CSTB2021TIAD-KPX0074)

中国医学装备

1672-8270

访问量0
|
下载量0
段落导航相关论文