基于模体的乳腺X线摄影质量控制中SNR变异系数的分析

沈茜刚 柴青焕 姜婷婷 沈悦 肖勤 顾雅佳

中国癌症杂志2025,Vol.35Issue(8):784-791,8.
中国癌症杂志2025,Vol.35Issue(8):784-791,8.DOI:10.19401/j.cnki.1007-3639.2025.08.007

基于模体的乳腺X线摄影质量控制中SNR变异系数的分析

Analysis of variation coefficient of SNR in phantom-based mammography quality control

沈茜刚 1柴青焕 1姜婷婷 1沈悦 1肖勤 1顾雅佳1

作者信息

  • 1. 复旦大学附属肿瘤医院放射诊断科,复旦大学上海医学院肿瘤学系,上海 200032
  • 折叠

摘要

Abstract

Background and purpose:In the quality control of mammography,the signal-to-noise ratio(SNR)refers to the ratio of the useful signal intensity to the background noise in the image,which is one of the important indicators for measuring the quality of the image.The coefficient of variation(CoV)is a commonly used indicator to describe the consistency and repeatability of SNR.This study aimed to assess the stability and repeatability of mammographic device performance by analyzing the changes in SNR CoV in two-dimensional(2D)images and tomosynthesis images(referred to as Tomo images)under different exposure modes using three mammographic devices from different manufacturers.Methods:A polymethylmethacrylate(PMMA)phantom designed for mammography quality control was used to perform automatic exposure detection at PMMA thicknesses ranging from 20-80 mm,with actual compression thickness equivalent to the average density of the breast compressed to 21-103 mm under full-field digital mammography(FFDM),low-dose mammography and digital breast tomosynthesis(DBT)exposure modes.The CoV of SNR in 2D images and tomosynthesis images was calculated for different mammographic devices under different exposure modes and compression thicknesses.Results:Between the compression thicknesses equivalent to the average density of the breast from 21 mm to 103 mm under FFDM,low-dose mammography,and DBT exposure modes,the differences in SNR CoV of 2D images under different exposure modes among mammographic devices 1,2,and 3 were statistically significant only in the DBT exposure mode(P=0.003),with SNR CoV ranging from 0.188%to 0.720%,0.368%to 1.073%and 0.402%to 1.662%,respectively.There were no statistically significant differences in SNR CoV of 2D images among devices 1,2,and 3 under FFDM and low-dose exposure modes(P=0.060).Under the DBT exposure mode,there were no statistically significant differences in the SNR CoV of the first projection image and the 0° projection image of tomosynthesis among devices 1(2 angles),2,and 3(P=0.373,P=0.742,P=0.225,P=0.693,respectively).Conclusion:The SNR CoV in 2D images and tomosynthesis images varies under different mammographic devices and exposure modes,with no fixed or standard values,but all within the required range for mammographic device quality control.The stability and repeatability of 2D images of mammographic devices are better under FFDM and low-dose exposure modes;the SNR CoV values of the first projection image and the 0° projection image of tomosynthesis under the DBT exposure mode show no statistical differences,indicating good stability of the devices.

关键词

乳腺X线摄影/信噪比/变异系数/质量控制

Key words

Mammography/Signal-to-noise ratio/Coefficient of variation/Quality control

分类

医药卫生

引用本文复制引用

沈茜刚,柴青焕,姜婷婷,沈悦,肖勤,顾雅佳..基于模体的乳腺X线摄影质量控制中SNR变异系数的分析[J].中国癌症杂志,2025,35(8):784-791,8.

基金项目

吴阶平医学基金会临床科研专项资助基金(320.6750.2022-11-24). Wu Jieping Medical Foundation Clinical Research Special Fund(320.6750.2022-11-24). (320.6750.2022-11-24)