首页|期刊导航|中国医疗设备|高BMI人群CCTA中P3T个性化对比剂方案与标准方案的图像质量及对比剂用量对比研究

高BMI人群CCTA中P3T个性化对比剂方案与标准方案的图像质量及对比剂用量对比研究

王宇鹏 张立新 吴柯薇 袁颖 李妍 胡艳军

中国医疗设备2025,Vol.40Issue(11):133-137,149,6.
中国医疗设备2025,Vol.40Issue(11):133-137,149,6.DOI:10.3969/j.issn.1674-1633.20251297

高BMI人群CCTA中P3T个性化对比剂方案与标准方案的图像质量及对比剂用量对比研究

Comparative Study on Image Quality and Contrast Agent Dosage Between P3T Personalized Contrast Agent Protocol and Standard Protocol in CCTA for High BMI Population

王宇鹏 1张立新 1吴柯薇 1袁颖 1李妍 1胡艳军1

作者信息

  • 1. 首都医科大学附属北京友谊医院 放射科,北京 100050
  • 折叠

摘要

Abstract

Objective To investigate the clinical application value of personalized patient protocol technology(P3T)in coronary CT angiography(CCTA)for populations with high body mass index(BMI)(28-32 kg/m2).Methods A total of 65 patients scheduled for CCTA scanning in Capital Medical University Affiliated Beijing Friendship Hospital from June to December 2023 were prospectively enrolled,and randomly divided into a control group(n=33)and a P3T group(n=32).The control group selected the corresponding contrast agent injection flow rate according to different BMI values.The P3T group generated personalized injection protocols based on patients'body weight through the P3T software.The two groups were compared in terms of contrast agent dosage,injection flow rate,effective dose(ED),results of subjective evaluation of image quality,and objective evaluation of image quality[including CT values of the aortic root and each segment of the coronary artery,signal to noise ratio(SNR),and contrast to noise ratio(CNR)].Results The contrast agent dosage and injection flow rate in the P3T group were significantly lower than those in the control group(P<0.05).There were no statistically significant differences between the two groups in CT values of the aortic root and each segment of the coronary artery,SNR,CNR,and ED(P>0.05).The subjective scores of images in both groups were≥3 points,which could meet diagnostic needs,with good consistency(Kappa=0.81).Conclusion The application of P3T technology in high BMI populations can reduce contrast agent dosage and injection flow rate while ensuring image quality,realize the personalization of injection protocols and consistency of examinations,and is clinically feasible.

关键词

个体化患者对比剂注射技术(P3T)/冠状动脉CT血管造影(CCTA)/体质量指数(BMI)/碘对比剂/图像质量/对比剂流速

Key words

personalized patient protocol technology(P3T)/coronary CT angiography(CCTA)/body mass index(BMI)/iodinated contrast agent/image quality/contrast agent flow rate

分类

特种医学

引用本文复制引用

王宇鹏,张立新,吴柯薇,袁颖,李妍,胡艳军..高BMI人群CCTA中P3T个性化对比剂方案与标准方案的图像质量及对比剂用量对比研究[J].中国医疗设备,2025,40(11):133-137,149,6.

基金项目

国家自然科学基金面上项目(82272068). (82272068)

中国医疗设备