首页|期刊导航|中国口腔种植学杂志|SmartX不同扫描策略在全口种植修复数字印模中的精度比较研究

SmartX不同扫描策略在全口种植修复数字印模中的精度比较研究

Gu Yanyan Wang Yuzhuo Fu Yuanfei Wang Jie

中国口腔种植学杂志2025,Vol.30Issue(6):574-579,6.
中国口腔种植学杂志2025,Vol.30Issue(6):574-579,6.DOI:10.12337/zgkqzzxzz.2025.12.010

SmartX不同扫描策略在全口种植修复数字印模中的精度比较研究

Comparative study on the accuracy of different SmartX scanning strategies in full-arch implant prosthetic digital impressions

Gu Yanyan 1Wang Yuzhuo 1Fu Yuanfei 1Wang Jie1

作者信息

  • 1. Department of Dental Technology,Shanghai Ninth People's Hospital,Shanghai Jiao Tong University School of Medicine&College of Stomatology&Shanghai Jiao Tong University&National Center for Stomatology&National Clinical Research Center for Oral Diseases&Shanghai Key Laboratory of Stomatology&Shanghai Research Institute of Stomatology,Shanghai 200011,China
  • 折叠

摘要

Abstract

Objective To evaluate the trueness of SmartX intraoral scanning(IOS)for full-arch digital implant impressions and to compare the impact of different scanning strategies on accuracy.Methods Scan bodies were arranged in two configurations(along-the-ridge and centripetal)on three all-on-6 maxillary models.SmartX IOS performed 60 scans(3 models×2 strategies×10 repetitions).Desktop scanner data served as a reference.IOS scans were compared against reference scans,assessing trueness and precision via distance/angular deviations between implant abutments and root mean square(RMS)errors.Results For along-the-ridge scans:distance deviation=75.25(45.35,136.40)μm,angular deviation=0.47°±0.22°,RMS=88.05(67.75,99.88)μm.For centripetal scans:distance deviation=113.90(81.93,315.60)μm,angular deviation=0.27°±0.18°,RMS=93.75(73.58,98.85)μm.Mann-Whitney U and paired-samples t-tests indicated significant precision differences between strategies(P<0.05).Conclusion Both SmartX strategies produced clinically acceptable deviations below passive fit thresholds for full-arch digital impressions.Scanning strategy is one of the key factors affecting the impression precision of the system.

关键词

全口种植/口内扫描/真实度/精度/体外实验

Key words

Full-arch implant/Intraoral scanning/Trueness/Precision/In vitro study

引用本文复制引用

Gu Yanyan,Wang Yuzhuo,Fu Yuanfei,Wang Jie..SmartX不同扫描策略在全口种植修复数字印模中的精度比较研究[J].中国口腔种植学杂志,2025,30(6):574-579,6.

基金项目

上海市卫健委青年项目(20224Y0356) Shanghai Municipal Health Commission Youth Research Program(20224Y0356) (20224Y0356)

中国口腔种植学杂志

1007-3957

访问量0
|
下载量0
段落导航相关论文