Abstract
A medical stainless steel cutting anastomat anvil was broken during the operation.The causes of fracture were analyzed by means of macroscopic observation,macroscopic examination,chemical composition analysis,scanning electron microscopy and energy spectrum analysis,mechanical property test and metallographic examination.The results show that the fracture mode of the anvil was intergranular brittle fracture,and the improper tempering temperature control was the root cause of the fracture of the anvil.The first type of temper brittleness was produced during the tempering process,and serious segregation occurred at the grain boundaries,which eventually led to a significant decrease in the impact toughness of the material.关键词
不锈钢/沿晶脆断/第一类回火脆性/偏析Key words
stainless steel/intergranular brittle fracture/the first type of temper brittleness/segregation分类
矿业与冶金引用本文复制引用
万天,钟世标,曾志卫,张文彬,吴冰冰..30Cr13不锈钢切割吻合器抵钉座断裂原因[J].理化检验-物理分册,2026,62(1):80-83,87,5.