首页|期刊导航|组织工程与重建外科杂志|小型猪下颌角骨折外固定装置力学性能的三维有限元研究

小型猪下颌角骨折外固定装置力学性能的三维有限元研究

张瑞娟 武江 高全文

组织工程与重建外科杂志2026,Vol.22Issue(1):34-41,8.
组织工程与重建外科杂志2026,Vol.22Issue(1):34-41,8.DOI:10.3969/j.issn.1673-0364.2026.01.004

小型猪下颌角骨折外固定装置力学性能的三维有限元研究

Three-dimensional finite element study of the mechanical properties of external fixation device for mandibular angle fracture in a miniature pig

张瑞娟 1武江 2高全文2

作者信息

  • 1. 100079 北京市 北京丰台康复医院
  • 2. 100048 北京市 解放军总医院第四医学中心烧伤整形医学部
  • 折叠

摘要

Abstract

Objective To construct a finite element model of a miniature pig's mandibular angle fracture with simulated installation of a self-developed external fixation device at the fracture site,and to analyze the device's biomechanical performance by three-dimensional finite element technology,thereby laying a foundation for clinical application.Methods A healthy Bama miniature pig weighing 20 kg was selected for cranial CT scanning.The scanned DICOM data were imported into Mimics for cranial three-dimensional reconstruction.The STL file was then imported into Geomagic Studio software to complete geometric reverse modeling.The resulting IGES file,a universal surface geometry file,was imported into Solidworks software to set up a virtual fracture at the mandibular angle.At the same time,the data of the self-developed mandibular external fixation device was imported into the Solidworks software for three-dimensional reconstruction,and a simulation installation was carried out at the fracture site of the mandibular angle.The IGES data of the mandible and the data of the external fixation device were imported into ICEM software for meshing to generate an UNS mesh file,which was then imported into Workbench software to analyze the stress and strain changes of the mandible and the external fixation device under a certain load.Results Three-dimensional finite element analysis revealed that after applying a vertically downward 196 N load at the first molar fossa of the right mandible,the maximum principal stress generated on both sides of the mandibular angle fracture ends was less than the ultimate strength of the mandible,103 MPa.The location of maximum Von-Mises stress of the screw was above the posterior end of the fracture,approximately 244.24 MPa.The location of maximum Von-Mises stress of the nut was also above the posterior end of the fracture,approximately 107.38 MPa.The location of maximum Von-Mises stress of the plate was at the screw hole below the anterior fracture end,approximately 138.54 MPa.The screw material was medical stainless steel 00Cr18ni14mo3,with an ultimate strength of 860 MPa.The nut and plate material was titanium alloy Tc4,with an ultimate strength of 895 MPa.The stresses produced by the three structures of the external fixation device were all below the ultimate strength of its material.The overall deformation of the external fixation device was 0.861 mm,and the strain was small.Conclusion The models of the mandibular angle fracture and the external fixation device of the miniature pig established using medical digital imaging software and three-dimensional finite element software have high simulation accuracy and can provide multiple simulation experiments.According to the results of the three-dimensional finite element analysis,the self-developed external fixation device for the mandible can achieve rigid fixation of the miniature pig's mandibular angle fracture.

关键词

外固定装置/下颌角骨折/三维有限元分析

Key words

External fixation device/Mandibular angle fracture/Finite element analysis

分类

医药卫生

引用本文复制引用

张瑞娟,武江,高全文..小型猪下颌角骨折外固定装置力学性能的三维有限元研究[J].组织工程与重建外科杂志,2026,22(1):34-41,8.