| 注册
首页|期刊导航|国际医学放射学杂志|60kVp管电压联合心脏深度学习重建在小儿先天性心脏病CTA中的应用

60kVp管电压联合心脏深度学习重建在小儿先天性心脏病CTA中的应用

孙明华 张飞飞 苏涛 潘玉坤 彭俪颖 张继良 谢瑞刚 葛英辉

国际医学放射学杂志2026,Vol.49Issue(2):137-143,7.
国际医学放射学杂志2026,Vol.49Issue(2):137-143,7.DOI:10.19300/j.2026.L22638

60kVp管电压联合心脏深度学习重建在小儿先天性心脏病CTA中的应用

Application of 60 kVp tube voltage combined with a cardiac deep learning reconstruction in CT angiography for pediatric congenital heart disease

孙明华 1张飞飞 2苏涛 2潘玉坤 2彭俪颖 3张继良 2谢瑞刚 2葛英辉1

作者信息

  • 1. 阜外华中心血管病医院放射科,郑州 451460||河南省心脏病影像医学重点实验室||河南省心血管疾病临床医学研究中心
  • 2. 阜外华中心血管病医院放射科,郑州 451460||河南省心脏病影像医学重点实验室
  • 3. 上海联影医疗科技股份有限公司
  • 折叠

摘要

Abstract

Objective To investigate the feasibility of 60 kVp CT angiography(CTA)combined with a cardiac deep learning reconstruction algorithm(CardioBoost)for children with congenital heart disease(CHD).Methods A total of 160 children with clinically confirmed CHD who were scheduled for preoperative cardiac CTA were prospectively enrolled,with a median age of 8 months.The children were randomly assigned to a conventional-dose group and a low-dose group(80 patients in each group).In the conventional-dose group,hybrid iterative reconstruction(HIR)was used to generate Group A images(80 kVp,400 mA,HIR).In the low-dose group,HIR and CardioBoost reconstruction were used to produce Group B1 images(60 kVp,240 mA,HIR)and Group B2 images(60 kVp,240 mA,CardioBoost reconstruction),respectively.Radiation doses between the standard-dose and the low-dose groups were compared,including volume CT dose index(CTDIvol),dose-length product(DLP),effective dose(ED),and size-specific dose estimate(SSDE).The subjective image quality scores(5-point scale)and objective image quality parameters of each cardiac chamber and great vessel,including image noise,signal-to-noise ratio(SNR),and contrast-to-noise ratio(CNR),were compared among Groups A,B1,and B2.Surgical findings were used as the gold standard to evaluate diagnostic accuracy.Quantitative data were compared using the Mann-Whitney U test and Kruskal-Wallis test,and categorical variables were compared using the chi-square test.Interobserver agreement between the two readers was assessed using Cohen's Kappa test.Results The ED in the low-dose group was reduced by 81.3%compared with that in the conventional-dose group(P<0.05).The interobserver agreement for the subjective image quality scoring was good across all three groups(all κ≥0.78).The subjective image quality scores were all≥3 in Groups A,B1,and B2.The subjective image quality score in Group B2 was significantly higher than that in Group B1(P<0.05),but did not differ significantly from that in Group A(P>0.05).Compared with Group B1,Group B2 showed reduced image noise and increased SNR and CNR(all P<0.05),and the objective image quality indices were comparable to or better than those of Group A.The diagnostic accuracy for intracardiac malformations in Group B2 was significantly higher than that in Group B1(P<0.05),but did not differ significantly from that in Group A(P>0.05).Conclusion For CTA in children with CHD,60 kVp combined with the CardioBoost algorithm can significantly reduce radiation dose while maintaining image quality comparable to that of the conventional-dose protocol.

关键词

低剂量/先天性心脏病/深度学习/CT血管成像/体层摄影术,X线计算机

Key words

Low-dose/Congenital heart disease/Deep learning/CT angiography/Tomography,X-ray computed

分类

医药卫生

引用本文复制引用

孙明华,张飞飞,苏涛,潘玉坤,彭俪颖,张继良,谢瑞刚,葛英辉..60kVp管电压联合心脏深度学习重建在小儿先天性心脏病CTA中的应用[J].国际医学放射学杂志,2026,49(2):137-143,7.

基金项目

河南省医学科技攻关计划项目(LHGJ20220123) (LHGJ20220123)

先进医用材料与医疗器械全国重点实验室项目(YGSKL-SHTech-2025-KF01) (YGSKL-SHTech-2025-KF01)

国际医学放射学杂志

1674-1897

访问量0
|
下载量0
段落导航相关论文