基于力学参数的血管介入机器人血管角度实时预测与安全预警策略初步研究

贾远望 陈韵岱 屈铭 辛然 刘子暖 杨诗怡 康文 王蔚然 刘肖 杨俊杰

上海交通大学学报(医学版)2026,Vol.46Issue(3):265-274,10.
上海交通大学学报(医学版)2026,Vol.46Issue(3):265-274,10.DOI:10.3969/j.issn.1674-8115.2026.03.001

基于力学参数的血管介入机器人血管角度实时预测与安全预警策略初步研究

Preliminary study on a mechanics-parameter-based strategy for real-time vascular angle prediction and safety warning in vascular interventional robots

贾远望 1陈韵岱 2屈铭 3辛然 4刘子暖 2杨诗怡 5康文 3王蔚然 2刘肖 5杨俊杰2

作者信息

  • 1. 中国人民解放军总医院第六医学中心心血管病医学部,北京 100048||中国人民解放军总医院研究生院,北京 100853
  • 2. 中国人民解放军总医院第六医学中心心血管病医学部,北京 100048
  • 3. 北京理工大学数学与统计学院,北京 100081
  • 4. 南开大学医学院,天津 300071||中国人民解放军总医院第六医学中心心血管病医学部,北京 100048
  • 5. 北京航空航天大学生物与医学工程学院,北京 100083
  • 折叠

摘要

Abstract

Objective·This study addresses the challenges of multimodal perception deficiencies and X-ray radiation dependency in vascular interventional robots.It investigated the mechanical characteristics of guidewire-vessel wall contact forces across varying vascular bending angles and established a contact force-angle mapping model to develop a novel robotic-assisted strategy integrating real-time vascular angle prediction and safety warning based on mechanical parameters.Methods·An in vitro vascular model(bending angles:0°‒80° at 5° intervals)was deployed on the R-One robotic platform.A force-sensing guidewire was advanced at 6 mm/s through curved segments.Using temporal registration,dynamic contact force variation trends(quantified as the rate of change)were extracted during the initial 2s traversal window.The correlation between vascular bending angles and variation rates was quantified.Based on this,a mapping model between contact force trends and vascular angles was constructed and evaluated using root mean squared error(RMSE)and mean absolute error(MAE).Results·A very strong positive correlation was observed between vascular bending angle and the rate of change in guidewire contact force during the initial phase of traversal through the bend(rs=0.98,P<0.001).This relationship exhibited distinct phases relative to the 0° baseline:the rate remained stable within the 0°‒25°,with no statistically significant differences;a significant increase first appeared at 30°(P.adj<0.05);beyond 50°,the rate of increase accelerated markedly,accompanied by a sharp enhancement in statistical significance(P.adj decreasing from 10-5 to 10-8).By 80°,the rate of change in contact force increased by 24,392.4%.Bayesian Information Criterion(BIC)-based changepoint analysis identified critical transition points at 35.60° and 50.65°,which closely align with the empirical thresholds of 30° and 50°,further confirming the structural nature of the relationship.The contact force-angle mapping model developed based on this relationship demonstrated excellent performance(R2=0.96,RMSE=4.93°,MAE=3.73°),significantly outperforming a conventional linear model(R2=0.89,RMSE=7.66°,MAE=5.03°).Conclusion·In the in vitro model,the rate of change of contact force during the initial phase of guidewire entry into a curved segment exhibited a significant positive correlation with vascular bending angle,characterized by distinct phase transitions and critical inflection points.Based on this feature,a contact force-angle mapping model was established,which outperformed the traditional linear model.This study preliminarily validates the feasibility of inferring vascular anatomical structures from mechanical characteristics,providing a theoretical basis for mechanics-based safety warning and auxiliary navigation.

关键词

血管介入机器人/力反馈/接触力变化速率/血管角度/安全预警与导航

Key words

vascular interventional robot/force feedback/rate of change of contact force/vascular curvature/safety warning and navigation

分类

医药卫生

引用本文复制引用

贾远望,陈韵岱,屈铭,辛然,刘子暖,杨诗怡,康文,王蔚然,刘肖,杨俊杰..基于力学参数的血管介入机器人血管角度实时预测与安全预警策略初步研究[J].上海交通大学学报(医学版),2026,46(3):265-274,10.

基金项目

国家自然科学基金(82470342,12361141817,62273046) (82470342,12361141817,62273046)

北京市科技新星计划(20230484471). National Natural Science Foundation of China(82470342,12361141817,62273046) (20230484471)

Beijing Nova Program(20230484471). (20230484471)

上海交通大学学报(医学版)

1674-8115

访问量0
|
下载量0
段落导航相关论文