同位素2026,Vol.39Issue(2):107-113,7.DOI:10.7538/tws.2025.youxian.073
基于cosRMC的BNCT剂量计算功能开发
Development of BNCT Dose Calculation Function Based on cosRMC
王胜哲 1姜韦 1杨训武 2卢棚 1张旭磊 2梁立振1
作者信息
- 1. 合肥综合性国家科学中心能源研究院(安徽省能源实验室),合肥 230031
- 2. 安徽理工大学 计算机科学与工程学院,淮南 232001
- 折叠
摘要
Abstract
To ensure the effectiveness and safety of boron neutron capture therapy(BNCT),precise dose calculation and treatment planning are of critical importance.In this study,the domestic and independent Monte Carlo particle transport code cosRMC was selected for dose calculation,and an automatic modeling and conversion process was developed to transform medical CT images into cosRMC input files.A publicly available head CT dataset was chosen as a case study.Through processes such as CT image processing,cosRMC modeling,neutron transport calculation,and dose calculation,the results were compared with those obtained from the OpenMC code.The findings indicated that the average dose error between cosRMC and OpenMC was less than 3%,preliminarily demonstrating that cosRMC can be used as a dose calculation tool in BNCT-TPS.关键词
硼中子俘获治疗/蒙特卡罗/剂量计算/cosRMCKey words
boron neutron capture therapy/Monte Carlo/dose calculation/cosRMC分类
能源科技引用本文复制引用
王胜哲,姜韦,杨训武,卢棚,张旭磊,梁立振..基于cosRMC的BNCT剂量计算功能开发[J].同位素,2026,39(2):107-113,7.