摘要
Abstract
[Objective]To analyze the typical faults(image artifacts and tube rotation errors)of the GE Discovery CT,propose a preventive maintenance strategy based on fault analysis,evaluate its implementation effect,and provide a practical basis for the efficient operation and maintenance of high-end CT equipment.[Methods]The troubleshooting and repair processes of the two types of faults were analyzed,based on which preventive maintenance strategies for the detector and the X-ray tube were formulated respectively.A pre-post comparative design was adopted,in which equipment operation data from 2023(pre-implementation)and 2024(post-implementation)were collected and analyzed.The changes in monthly average fault frequency,maintenance cost,downtime,and patient satisfaction were compared,and paired t-tests were used for statistical analysis.[Results]After the implementation of the preventive maintenance strategy,the monthly average fault frequency decreased by 66.4%,the maintenance cost decreased by 58.0%,the downtime was reduced by 69.5%,and patient satisfaction increased by 15.9%.All differences were statistically significant(P<0.05).[Conclusion]The preventive maintenance strategy developed based on typical faults can significantly reduce the fault frequency and maintenance cost of the GE Discovery CT,shorten its downtime,and improve patient satisfaction.It provides a quantifiable and generalizable practical solution for the scientific operation and maintenance of high-end CT equipment.关键词
GE Discovery CT/故障分析/预防性维护/医疗设备管理Key words
GE Discovery CT/fault analysis/preventive maintenance/medical equipment management分类
医药卫生