生物医学工程研究2026,Vol.45Issue(3):221-226,6.DOI:10.19529/j.cnki.1672-6278.2026.03.06
强迫振荡肺功能测试的呼吸电抗校准方法
Method for calibrating respiratory reactance based on the forced oscillation pulmonary function test
摘要
Abstract
Forced oscillation technique(FOT)is a critical tool in pulmonary function testing.Aiming at the lack of a unified standard for electrical reactance in existing FOT equipment,which leads to significant differences in the electrical reactance results of different instruments,we proposed an easily reproducible design method for reactance standards that covers the human physiological range selected commercial FOT devices(e.g.,Canadian Thorasys tremoflo® C2,German Jaeger IOS)to test these standards and recor-ded absolute errors for each instrument.Furthermore,5 healthy subjects were recruited for validation tests.The results showed that cali-bration based on theoretical values significantly reduced the individual differences among the instruments for the same subjects,with an average reduction of 51.4%in standard deviation during the test.This method can lay a foundation for cross-device data interoperability and clinical standardization.关键词
呼吸电抗/校准方法/强迫振荡技术/肺功能测试/傅里叶变换/电抗标准物Key words
Respiratory reactance/Calibration method/Forced oscillation technique/Pulmonary function test/Fourier transform/Reactance standards分类
医药卫生引用本文复制引用