摘要

Abstract

Objective To explore the safety and efficacy of the single-operator controllable reduction device for distal ra-dius fractures in clinical application.Methods A retrospective analysis was conducted on 21 patients with extension-type distal radius fractures admitted from September 2023 to September 2024,including 6 males and 15 females.Their ages ranged from 26 to 84 years old,with a mean age of(62.09±13.43)years old.All patients received reduction and fixation treatment using the single-operator reduction device.The differences in distal radioulnar variance,ulnar inclination angle and palmar tilt angle of patients were compared before and after reduction.All measurements were performed by two independent physicians,and the average values were adopted.Twelve weeks after conservative treatment,wrist joint function was evaluated using the Cooney wrist scoring system.Results All patients achieved clinical fracture healing and were followed up for more than 12 weeks.Be-fore reduction,the distal radioulnar variance was(3.071±2.719)mm,ulnar inclination angle was(19.048±5.133)°,and pal-mar tilt angle was(-18.333±12.220)°.After reduction,the above indicators were(1.252±2.540)mm,(21.476±3.907)° and(5.714±7.478)° respectively.There were statistically significant differences in all indexes before and after reduction(P<0.05).At 12 weeks after reduction,the mean Cooney wrist joint functional score was(89.04±4.85)points,including 12 cases with ex-cellent results,6 good cases and 3 fair cases.Conclusion The single-operator controlled reduction device provides reliable ra-diographic restoration for Colles-type distal radius fractures,with favorable short-term functional recovery outcomes.

关键词

桡骨远端骨折/复位装置/临床应用

Key words

Distal radius fracture/Reduction device/Clinical application

分类

医药卫生

引用本文复制引用

施纯南,王志杰,许永权,吴鸿斌,许培阳,陈志钦,张金山..单人可控桡骨远端骨折复位装置的早期临床应用[J].中国骨伤,2026,39(7):660-664,5.

基金项目

泉州市科技计划项目(编号:2023NS101)Quanzhou City Science and Technology Program(No.2023Ns101) (编号:2023NS101)

中国骨伤

1003-0034

访问量0
|
下载量0
段落导航相关论文