中国内镜杂志2026,Vol.32Issue(8):86-90,5.DOI:10.12235/E20260071
不同规格支气管镜管道干燥时间探讨
Study on air-inflation duration for complete drying of the lumen of bronchoscopes with different specifications
李敏 1王缃兰 1吴熳 1葛姝彤 1武晨东 1许锋1
作者信息
- 1. 武汉大学人民医院 内镜中心,湖北 武汉 430060
- 折叠
摘要
Abstract
Objective This study aimed to investigate the required air-injection duration for drying bronchoscope channels of different specifications.Methods Two types of bronchoscopes were manually cleaned and disinfected after clinical use.A pressure air gun at 0.4 MPa was used to inject air into the biopsy channel and the suction channel separately.After air injection,cobalt chloride test paper was fully contacted with the inner lumen,and drying was determined based on the color change of the paper.In the first group,air was injected into both channels for 10 seconds,with 30 repeated tests.The injection time was then increased by 10 seconds per group until all 30 tests indicated dry channels.Results For a bronchoscope with a biopsy channel diameter of 2.0 mm,air injection for 30 seconds into each channel achieved drying.For a bronchoscope with a biopsy channel diameter of 3.0 mm,air injection for 40 seconds was required.Conclusion Different air-injection times are recommended for bronchoscopes of different specifications:30 seconds for those with a 2.0 mm biopsy channel and 40 seconds for those with a 3.0 mm biopsy channel.关键词
支气管镜/管道直径/干燥/注气时间/氯化钴试纸/吸引管道/活检管道Key words
bronchoscope/channel diameter/drying/air-injection time/cobalt chloride test paper/suction channel/biopsy channel分类
医药卫生引用本文复制引用