双能CT迭代去金属伪影重建算法联合虚拟单能成像在腰椎金属内固定伪影抑制中的应用及能量优化

聂泽花 田惠民 包静文 王荣德 李晓丹

实用临床医药杂志2026,Vol.30Issue(16):32-39,8.
✕
实用临床医药杂志2026,Vol.30Issue(16):32-39,8.DOI:10.7619/jcmp.20262624

双能CT迭代去金属伪影重建算法联合虚拟单能成像在腰椎金属内固定伪影抑制中的应用及能量优化

Application and energy optimization of dual-energy CT iterative metal artifact reduction reconstruction algorithm combined with virtual monochromatic imaging for metal artifact suppression in lumbar spinal instrumentation

聂泽花 1田惠民 2包静文 1王荣德 1李晓丹2

作者信息

  • 1. 河西学院医学院影像教研室,甘肃张掖,734000
  • 2. 张掖市第二人民医院影像中心,甘肃张掖,734000
  • 折叠

摘要

Abstract

Objective To investigate the application of dual-energy CT iterative metal artifact re-duction reconstruction algorithm(DE-iMAR)combined with virtual monochromatic imaging(DE Mono+)for metal artifact suppression in lumbar spinal instrumentation and to identify the optimal reconstruction energy level.Methods This retrospective study enrolled 105 patients who underwent CT examina-tion after lumbar spinal instrumentation.All patients were scanned using a third-generation dual-source CT scanner(Siemens SOMATOM).The original dual-energy CT mixed images were acquired(group A),and three reconstruction approaches were applied:DE-iMAR alone(group B),DE Mono+alone(group C),and DE-iMAR combined with DE Mono+(group D).Two radiologists with more than 5 years of experience independently evaluated image quality and artifact scores using a double-blind method.The subjective scores,CT values of the psoas major muscle at the most se-vere artifact level and at artifact-free level were compared among the four groups under different en-ergy conditions.The inter-observer agreement for image quality and artifact scoring was assessed.The subjective scores,artifact scores,subjective scores at the most severe artifact site,and artifact index(AI)values were compared among the four groups.Results At 130 keV,the subjective scores of groups C and D were significantly higher than those at 120 kVp and at 40,70,100,160,and 190 keV(P<0.05).At 40-190 keV,group D had significantly higher subjective scores than group C at the same energy levels(P<0.05);at 120 kVp,group D had significantly higher sub-jective scores than groups A and B(P<0.05).At 120 kVp,the CT values of the psoas major mus-cle at the most severe artifact level in groups A and B were significantly higher than those at the arti-fact-free level(P<0.05).At 120 kVp and at 40,70,100,160,and 190 keV,the CT values at the most severe artifact level in groups C and D were significantly higher than those at the artifact-free level(P<0.05).At 40-190 keV,the CT values of the psoas major muscle at both the most severe artifact level and the artifact-free level in group D were significantly lower than those in group C(P<0.05).At 120 kVp,the CT values at both the most severe artifact level and the artifact-free level in group D were significantly lower than those in groups A,B,and C(P<0.05).At 130 keV,groups C and D showed the optimal image quality and artifact suppression;therefore,subsequent analyses were performed at this energy level.The intraclass correlation coefficients(ICC)for image quality and artifact scoring by the two radiologists were all over 0.8,indicating good agreement(P<0.05).The subjective scores of group D were significantly higher than those of groups A,B and C,with groups B and C scoring higher than group A(P<0.05).The artifact scores of group D were significantly lower than those of groups A,B,and C,with groups B and C scoring lower than group A(P<0.05).The subjective scores of group D for soft tissue,metal-bone interface,and e-rector spinae were significantly higher than those of groups A,B,and C,with groups B and C sco-ring higher than group A(P<0.05).The AI values of region of interest(ROI)1,ROI 2,and ROI 3 in group D were significantly lower than those in groups A,B,and C,with groups B and C scoring lower than group A(P<0.05).Conclusion The combination of DE-iMAR and DE Mono+demonstrates application value in reducing metal artifacts in lumbar spinal instrumentation and significantly improves CT image quality after lumbar spinal instrumentation,providing more reli-able imaging evidence for clinical diagnosis.

关键词

双能CT/迭代去金属伪影重建算法/虚拟单能成像/金属伪影/腰椎内固定术/伪影指数/图像质量/重建能量

Key words

dual-energy CT/iterative metal artifact reduction reconstruction algorithm/virtual monochromatic imaging/metal artifacts/lumbar spinal instrumentation/artifact index/image quali-ty/reconstruction energy

分类

医药卫生

引用本文复制引用

聂泽花,田惠民,包静文,王荣德,李晓丹..双能CT迭代去金属伪影重建算法联合虚拟单能成像在腰椎金属内固定伪影抑制中的应用及能量优化[J].实用临床医药杂志,2026,30(16):32-39,8.

基金项目

国家级大学生创新创业训练项目(202210740052) (202210740052)

甘肃省教育厅高校教师创新基金项目(2026A-184) (2026A-184)

河西学院校长基金青年科研项目(QN2024009) (QN2024009)

实用临床医药杂志

1672-2353

访问量0
|
下载量0
段落导航相关论文