Feasibility study of using third-generation dual source dual-energy CT virtual non-contrast images instead of true non-contrast images to evaluate parotid gland lesions

张璐瑶 1郭方凯 1李宜泽 1黄晓颖 1陈英敏1

作者信息

  • 1. 河北省人民医院医学影像科,河北 石家庄 050051
  • 折叠

摘要

Abstract

Objective:To evaluate the feasibility of virtual non-contrast(VNC)images using dual source dual-energy CT instead of true non-contrast(TNC)images in patients with parotid gland lesions.Methods:Forty-six patients(with 54 parotid lesions)who underwent preoperative dual-energy enhanced CT examination of the parotid gland and pathologically confirmed parotid masses were prospectively included.VNC images in the arterial and venous phases(VNCa,VNCv)were reconstructed,and the consistency of CT values of parotid gland lesion,normal parotid tissue,submandibular gland,sternocleidomastoid muscle,subcutaneous fat,and internal jugular vein between the two groups of VNC images and TNC images was analyzed.The absolute attenuation error of CT value measurement results between the two groups of VNC images and TNC images(VNCaerror,VNCverror)was calculated.The image noise,lesion signal-to-noise ratio(SNR),contrast-to-noise ratio(CNR),subjec-tive image quality scores,and lesion detection rates were compared among the TNC,VNCa,and VNCv images.Results:Except for the sternocleidomastoid muscle,there were statistically significant differences in CT values of other tissues on VNCa(or VNCv)and TNC images(P<0.05),but the mean difference did not exceed 5 HU.The percentage of VNCaerror<15 HU in all measurements was 94%,and the percentage of VNCverror<15 HU was 87%.The Bland-Altman analysis showed a high consistency between the CT values on VNC images and TNC images.The lesion CNR of the two VNC images was lower than that of the TNC images(P<0.05),and there was no statistically significant difference in the noise and lesion SNR between the VNCa images and the TNC images.The overall subjective image quality scores of the three groups of images showed no statistically significant difference,while the lesion boundary and significance scores of the TNC images were higher than those of the two VNC images(P<0.05).The detection rate of parotid gland lesions in the three groups of images was the same.The radiation dose of parotid gland dual energy arterial and venous phase scanning was reduced by about 33.50%compared to conventional three-stage scanning.Conclusion:In the evaluation of parotid gland lesions,VNC images obtained from third-generation dual-source dual energy CT can reduce radiation dose while meeting imaging diagnostic requirements.Compared with VNCv images,VNCa can provide more reliable baseline CT values and higher image quality.

关键词

腮腺肿瘤/体层摄影术,X线计算机

Key words

Parotid Neoplasms/Tomography,X-Ray Computed

分类

医药卫生

引用本文复制引用

张璐瑶,郭方凯,李宜泽,黄晓颖,陈英敏..第3代双源CT虚拟平扫替代常规平扫评估腮腺病变的可行性[J].中国临床医学影像杂志,2026,37(9):624-628,639,6.

基金项目

2024年度河北省政府资助临床医学优秀人才培养项目(编号:2023.10-2026.9). (编号:2023.10-2026.9)

中国临床医学影像杂志

1008-1062

访问量0
|
下载量0
段落导航相关论文