基于脉冲高压消减电荷累积的四极杆质谱检测技术研究

孙澎勇 林一明 孟坚鑫 张敬 曹祥宽 沈林杰

分析化学2026,Vol.54Issue(9):1557-1564,中插21-中插22,10.
✕
分析化学2026,Vol.54Issue(9):1557-1564,中插21-中插22,10.DOI:10.19756/j.issn.0253-3820.261165

基于脉冲高压消减电荷累积的四极杆质谱检测技术研究

Research on Quadrupole Mass Spectrometry Analysis Technology Based on Pulsed High Voltage Charge Accumulation Mitigation Method

孙澎勇 1林一明 2孟坚鑫 1张敬 1曹祥宽 2沈林杰2

作者信息

  • 1. 杭州海康威视数字技术股份有限公司,杭州 310053
  • 2. 杭州海康威视元物智科技有限公司,杭州 310053
  • 折叠

摘要

Abstract

The stability of mass spectrometers is a key indicator,directly affecting the accuracy and reliability of experimental results.Nevertheless,charge accumulation on the fixed ceramic insulators of quadrupole alters the ion transmission efficiency and the full width at half maximum(FWHM)of mass spectral peaks,thus degrading instrument stability.To quantify the influence of charge accumulation on stability,simulation experiments were carried out.The simulation results revealed that charged ceramics caused obvious deflection of ion trajectories and distortion of the quadrupole field.Moreover,greater bias voltages applied to the ceramics led to more severe spectral peak variations.At a ceramic voltage of 20 V,the ion transmission efficiency decreased by 53.3%,the FWHM shifted by 0.23 amu,and the adverse effect intensified with the increase of mass-to-charge ratio.To solve this issue,the pulsed high voltage charge accumulation mitigation method was proposed.This method consisted of stabilization process,scanning process,and charge elimination process.In the elimination process,a DC high voltage of approximately 500 V amplitude and an RF high voltage with a peak-to-peak value of approximately 6000 V were applied to periodically eliminating charges.Two validation experiments,including FWHM drift monitoring and stability comparison of the mixed standard solution,were performed to verify the performance of the proposed method.In FWHM test,fragment ions of perfluorotributylamine were monitored under the conditions with and without the pulsed high voltage charge accumulation mitigation method.For the stability test,40 consecutive injections of a mixed solution containing hexachlorobenzene,hexabromobenzenen and decabromodiphenyl ether were tested with the two working modes.The average relative standard deviation of peak areas for the three analytes was reduced from 8.61%to 3.93%,which indicated that the proposed method remarkably improved the stability of the mass spectrometer.

关键词

质谱仪器稳定性/电荷累积/脉冲高压消减电荷累积方法/质谱峰形

Key words

Mass spectrometer stability/Charge accumulation/Pulse high voltage charge accumulation mitigation method/Mass spectral peak profiles

引用本文复制引用

孙澎勇,林一明,孟坚鑫,张敬,曹祥宽,沈林杰..基于脉冲高压消减电荷累积的四极杆质谱检测技术研究[J].分析化学,2026,54(9):1557-1564,中插21-中插22,10.

基金项目

杭州海康威视数字技术股份有限公司企业自主研发项目(No.PJ08PD20240115288)资助. Supported by the Independent R&D Project of Hangzhou Hikvision Digital Technology Co.,Ltd.(No.PJ08PD20240115288). (No.PJ08PD20240115288)

分析化学

0253-3820

访问量0
|
下载量0
段落导航相关论文