2个治疗计划系统调强放疗剂量验证比较OACSTPCD
Comparison of IMRT Dose Verification between Two Treatment Planning Systems
目的:通过比较Xio和Oncentra MasterPlan 2个治疗计划系统调强放疗剂量验证的结果,评价它们应用于临床的准确性.方法:用IBA二维电离室矩阵MatriXX分别采集Xio和Oncentra MasterPlan系统的鼻咽癌调强照射野数据,并用IBA Omnipro I'mRT1.6软件分析处理.结果:Xio系统62个照射野3 mm 3%通过率均值为95.4%,3 mm 5%为98.4%;Oncentra MasterPlan系统7…查看全部>>
Objective To value the clinical accuracy of Xio and Oncentra MasterPlan treatment planning systems for their IMRT dose verification. Methods Both the NPC IMRT beam dose data of Xio and Oncentra MasterPlan systems were collected by IBA MatriXX. Then they were analyzed by IBA OmniPro I'mRT 1.6 software. Results The 3mm 3% average pass rate of 62 beams in Xio system was 95.4% and the 3mm 5% was 98.4%. The 3mm 3% average pass rate of 75 beams in Oncentra M…查看全部>>
陈杰;朱云云;朱锦锋
南京军区福州总医院放疗科,福州350025南京军区福州总医院放疗科,福州350025南京军区福州总医院放疗科,福州350025
医药卫生
调强放疗剂量验证Gamma通过率XioOncentra MasterPlan
IMRT dose verification Gamma pass rate Xio Oncentra MasterPlan
《医疗卫生装备》 2012 (9)
111-112,2