基于PSNR和SSIM方法评估双能量CT肝脏虚拟平扫图像质量研究

Assessment of the Image Quality of Virtual Non-Contrast Dual-energy CT Liver Scans Using Both PSNR and SSIM Methods

李亚光 1李洁 2薛廷玉 1母建奎 3王强 3郑立冬 1王勇 1雷立存4

作者信息

  • 1. 河北医科大学第一医院放射与核医学科,石家庄 050000
  • 2. 河北医科大学医学影像学院,石家庄 050000
  • 3. 河北医科大学第一医院胸外科,石家庄 050000
  • 4. 河北医科大学第一医院放射与核医学科,石家庄 050000||河北医科大学医学影像学院,石家庄 050000
  • 折叠

摘要

Abstract

The purpose of this study was to investigate the feasibility of replacing true non-contrast(TNC)dual-energy computed tomography(DECT)images with virtual non-contrast(VNC)DECT images by comparing their quality on the basis of both the peak signal-to-noise ratio(PSNR)and structural similarity index(SSIM).Methods:A prospective analysis was conducted on TNC and enhanced three-phase DECT images of the livers of 33 patients.Post-processing was used to obtain the arterial-phase(VNCa),venous-phase VNC(VNCv),and delayed-phase VNC(VNCd)images.Both the PSNR and SSIM methods were used to compare the overall and local TNC and three-phase VNC images.The CT numbers and noise values(standard deviation)of the liver and erector spinae muscle were measured,and the SNR and contrast-to-noise ratio(CNR)were calculated.The dose length product values of the TNC and enhanced three-phase VNC scans were recorded,and the objective evaluation indicators and radiation doses of the three-phase VNC and TNC images were compared.Bland-Altman scatter plots were drawn to analyze the consistency of the liver CT numbers,SNRs,and CNRs.Results:The overall comparison of the three-phase VNC and TNC images showed PSNR values of(18.01±1.06),(18.33±0.99),and(18.20±1.04)and SSIM values of(0.76±0.04),(0.77±0.03),and(0.78±0.04),with the differences being not statistically significant.The local comparison of these images showed PSNR values of(29.90±2.50),(30.97±2.34),and(30.61±2.76)and SSIM values of(0.75±0.04),(0.77±0.03),and(0.77±0.04),and the differences were also not statistically significant.The CT number of the liver in the three-phase VNC image was higher than that in the TNC image.The CNR of the three-phase VNC image and the SNR of the VNCv image were not statistically different from those of the TNC image,and the liver CT numbers,SNRs,and CNRs in the three-phase VNC and TNC images were highly consistent.Using the VNC+three-phase enhancement scheme can reduce the radiation dose by approximately 29.63%by removing the TNC part.Conclusion:The VNC DECT image of the liver is of good quality and can accurately reproduce the TNC image,meeting clinical diagnostic needs.

关键词

双能量CT/虚拟平扫/结构相似性指数/峰值信噪比

Key words

dual energy computed tomography/virtual non-contrast scan/structural similarity index/peak signal to noise ratio