Objective:This study aimed to investigate the radiation dose levels and typical values of computed tomography(CT)devices in our unit,providing a basis for optimizing radiation protection.Methods:The scanning parameters and radiation dose data of head,chest,and abdominal examinations from five CT devices in our hospital from July to December 2023 were retrospectively analyzed.Median values were determined and compared with domestic and international diagnostic reference levels(DRL).Results:The typical values for the volume CT dose index(CTDIvol)and dose length product(DLP)were 54.40mGy and 735.80mGy·cm,14.20mGy and 443.60mGy·cm,and 16.40mGy and 678.20mGy·cm for head,chest,and abdomen,respsectively.The highest CTDIvol and DLP typical values were observed in head examinations,followed by abdominal and chest.CTDIvol and DLP for head examinations slightly increased with age.Significant statistical differences were found in CTDIvol and DLP for head and abdominal CT examinations with different years of use.Here,the DLP values were lower than those in other countries.Excluding a slightly higher head DLP,all other radiation doses were below the national health industry standards.Abdominal DLP was higher than other provinces and cities.Conclusion:Radiation doses of adult CT examinations in this survey are consistent with domestic and international results.Standardized training and optimization of the scanning range in CT examinations are needed to reduce unnecessary radiation exposure to patients.关键词
CT/辐射剂量/诊断参考水平Key words
CT/adiation dose/diagnostic reference levels分类
医药卫生引用本文复制引用
张斌,杜静波,张博洋,佟硕,魁学妍,冯陶然,杨海龙,廖建勇,张永县..成年人CT检查辐射剂量及其典型值调查与分析[J].CT理论与应用研究,2025,34(3):339-344,6.基金项目
北京市大兴区人民医院课题基金(大兴区人民医院CT检查辐射剂量调查与分析(4202430521)). (大兴区人民医院CT检查辐射剂量调查与分析(4202430521)